Laboratory - model making                         can be designed and used accordingly in the dental technology la-
                                                              boratory. At least four implants are required for immediate load-  bar system
            The technician again checks the exact fit of the respective impres-  ing.
            sion body in the impression material before selecting the appro-                                    Three versions are available for the laboratory production of bar
                                                                                                                constructions. The gold bar cylinder (18) is used by the dental tech-
            priate laboratory analogue (8 | 9) and fastening it to the impres-  MeoLock®
            sion post (5 | 7) with the laboratory screw (50). After the applica-                                nician for casting. The plastic bar cylinder (19) is embedded to-
            tion of a gingiva mask made of hardening rubber and the actual   Denture prosthesis is secured against vertical pull-off forces by the   gether with the modeled bar construction and cast in metal. This
            model production with plaster of paris, the intended abutments   MeoLock® abutment system . One-piece and two-piece MeoLock®   plastic cylinder can be burned out. Another bar cylinder is made of
            can be further processed.                         abutments (21 + 22) are available in heights of 1-6 mm. After ex-  titanium (20) and can be fixed with a laser in the laboratory. The
                                                              posure, the one-piece MeoLock® abutment (21) is screwed onto   selected bar abutments are fixed on the model with the laboratory
                                                              the implants (1 |1.1) with the MeoLock® insertion tool with 30   screw (50) and can now be processed. After completion of the bar
            Laboratory – abutment selection and processing
                                                              Ncm. The two-part MeoLock® abutment (22) is screwed onto the   construction, it is fastened in the mouth with the final connecting
                                                              implant (1 |1.1) with the connecting screw (51) with 30 Ncm and   screw (51) with 30Ncm.
            Depending on your wishes or the axis of the placed implant, abut-  provided with an impression cap (23) for the impression. After the
            ments (10 - 16 | 18 - 22 | 27 | 33 | 34 | 46 | 48 | 49) can be selected   impression has been taken, the dental technician places the Me-
            and individualized from the selection offered. Standard abutments   oLock® model analog (24) in the impression cap (23) remaining in   CLEANING AND STERILIZING INSTRUCTIONS
            (10 | 12) are sufficient if the implant is aligned correctly. For minor   the impression and fabricates the master model. The cover den-  All Meoplant implants are delivered sterile and are intended for
            corrections, the one-piece structure (13) can be adjusted to the   ture prosthesis is now constructed over the MeoLock® attachment   single use only before the stated expiry date.
            tooth axes by grinding in the mouth using a rubber dam. The uni-  (21 | 22). The MeoLock® matrix (25) can be polymerized both in
            versal abutments (15 | 16) can be individualized in the laboratory.   the mouth and on the model . The colors of the retention elements   Warning:  The  components that are delivered sterile must  no
                                                              (26) (patrix) symbolize the different pull-off forces. If the retention   longer be used if the packaging is damaged or has been previously
            open healing                                                                                        opened.
                                                              elements are worn , they can be replaced.
                                                                                                                INFORMATION ON MRI SAFETY
            suitable healing caps made of titanium (3) with 2-4 Ncm or custom-  ball head attachments
            izable healing caps made of Peek (4) with 15 Ncm can be screwed                                     In the case of magnetic resonance examinations, it should be
            onto the connecting screw (5) with open healing, depending on the   Ball head abutments (27) with the Insertion instrument (62) with   noted that the Meoplant implant system has not been tested for
            thickness of the gingiva, in order to close the peri-implant soft tis-  30 Ncm and with the MeoMini® implant (47) with 25 Ncm. A ball   safety and compatibility in MRI examinations.
            sue to form and mature. Impression as described under a) and b).   head model analogue (28) is available for model production. For
                                                              the impression, the ball head abutments (46) for MeoMini® (47)   HANDLING AND STORAGE
                                                              and the ball head abutments (27) for MeoClassic (1) and MeoTulip
                                                              (1.1) are screwed on and conventionally molded. After removing   The implants are gamma sterilized in the packaged condition and
            Open healing with abutments and long-term temporary   the impression tray, the ball head analogue (28) is inserted into the   are intended for single use. There must be no resterilization. The
                                                              resulting cavity and the master model is made in the laboratory.   implants may no longer be used if the packaging is damaged or
            If the soft tissue maturation is to proceed as undisturbed as possi-
            ble, frequent changing of structural parts is contraindicated. Here   With both types of implants , the dental technician has the option   opened and if the expiry date has expired. The manufacturer de-
            it makes sense to insert the abutments in the same way as the later   of individually fitting the two existing ball head abutments (27 |   clines all responsibility in the event of non-compliance. All instru-
            dentures. Depending on the fit of the implant, straight abutments   46) with two different retention abutments. Firstly, the rubber O-  ments and abutments/secondary parts are packaged non-sterile
            (10 | 11 | 13 - 16) can be used, which can be individually adjusted   ring (30) and the associated O-ring mount (29) as a matrix that is   and must be sterilized before each use (can be found at: https://re-
            to the tissue conditions by grinding .            polymerized into the prosthesis and secondly, the Preci - Clix male   processing.meoplant.de/ and as a paper version with the delivered
                                                              (29) with the associated Preci - Clix matrix ( 31) which is also pol-  product). For all items intended for single use, it was not tested
            Open healing with abutments and prefabricated dentures for im-  ymerized into the prosthesis .      whether repeated cleaning procedures and sterilization processes
            mediate loading                                                                                     have an effect on the function and fitting accuracy of the material.
                                                              MeoMini® bodies                                   Therefore, multiple use of this item is strongly discouraged. The
            The only difference from the previous paragraph is the supply. As                                   product must be stored in a dry place in its original packaging, at
            part of CAD/CAM planning, the dental technician can then digitally   The one-piece MeoMini® compact abutment (48) is screwed onto   room temperature and away from direct sunlight. Improper stor-
            scan the dentures using scan abutments (17 | 43), which are   the MeoMini® implant (47) in the mouth with the insertion instru-  age can affect product properties and lead to supply failure.
            screwed onto the implant that has just been placed in the mouth.   ment (60 | 61) with 25 Ncm and  then molded. The  two-part
            With the resulting data set, the crown, bridge or bar restoration   MeoMini abutment (49) is fixed to the MeoMini® implant (47) with
                                                              the connecting screw (53) with 25Ncm and then molded.