proper use of the surgical and prosthetic parts can result in dam-  implant system                  implant is designed in such a way that the implant edge should be
            age to the implant and the abutment parts, resulting in bone loss                                   flush with the crestal bone edge after insertion. The MeoTulip im-
            or even loss of the implant. Only properly trained dentists who are                                 plant is designed in such a way that the polished tulip-shaped area
            appropriately trained in implantology in relation to diagnosis, plan-                               should be visible above the crestal bone wheel after insertion. The
            ning and surgical technology may use the Meoplant® Medical Im-                                      diameter-reduced MeoMini® implant has an additional fanned, tu-
            plant System. The Meoplant® Medical Implant System is a coordi-                                     lip-shaped polished shape instead of an internal hexagon, which
            nated implant system that must not be combined with structures                                      merges into an external hex (Figure 2).
            and instruments from other implant manufacturers. In our terms
            of delivery, we guarantee the perfect condition and proper func-
            tioning of our products if all the above rules are observed and com-
            plied with. Before using the Meoplant® Medical implant system,
            the practitioner must ensure that all required product and system
            parts are in perfect condition and that they are secured against as-
            piration and swallowing during treatment. If there are any ques-
            tions regarding the use or the indication limits before the opera-
            tion, the operation must not be performed until these questions
            have been clarified. The practitioner concerned is solely responsi-
            ble for any damage caused by the use of Meoplant® Medical prod-                                            Figure 2: MeoMini • MeoClassic • MeoTulip - implant
            ucts. The abutments/secondary parts may only be used according
            to their indication in accordance with the general rules for dental                                 The Meoplant® Medical Implant System consists of implants of dif-
            and surgical activities, taking into account the occupational health                                ferent diameters and lengths intended for different indications. It
            and safety and accident prevention regulations. Diameter reduced                                    is expressly pointed out that compliance with the prescribed drill-
            implants and angled abutments are not suitable for use in the pos-  Figure 2: MeoClassic® and MeoMini® implant with closure screws   ing protocol is necessary. Separate  components such as locking
            terior region. Advice on the use of Meoplant products takes place                                   screws, gingiva formers, impression posts, MeoLock®, MeoMulti®
            regularly and is strongly recommended.            Meoplant® implants are made of grade 4 KV pure titanium. It con-  and Meoplant® prosthetic components are available for the Meo-
                                                              sists of endosseous implants of different sizes (see Tables 1 and 2)   plant® implant system. The MeoClassic and MeoTulip abutments /
            DESCRIPTION                                       and coordinated components, which are described below with re-  secondary parts are compatible with each other.
            The implants from Meoplant® Medical are delivered  sterile   gard to indication and application. The surface of the Meoplant®   Abutments and secondary prosthetics
            (gamma-sterilized) in double packaging. Before use, check the   implants are blasted and acid-etched (SLA). The apically confis-
            packaging for damage. The implants must be stored dry and at   cated implant body has a cylindrical, parallel -walled shape with a
            room temperature. If the expiry date has passed, these implants   macro-thread in the central part and a micro-thread in the crestal
            may no longer be used.                            area. The three cutting edges ensure a self-centering insertion. The
                                                              outer portions of the thread are designed so that a self-tapping ef-
                                                              fect occurs as long as the proposed drilling protocol is followed.
                                                              The thread pitch and the furrow-like interruption of the thread has
                                                              the function of collecting bone chips with vital bone cells, trans-
                                                              porting them to the implant body and compacting them in order
                                                              to biologically optimize the primary stability and healing into the
                                                              bone. The tip of the implant has the shape of a lens and its convex
                                                              area protects anatomically endangered structures such as the
                                                              maxillary sinus mucosa in a sinus lift. The main purpose of the hex-
                                                              agon socket (hex connection) is to ensure a torsionally stable, per-
                                                              manently congruent  connection between the  implant and the
                                                              abutment. The cone connection ensures an intimate metal con-
                                                              nection, which reduces the micro gap (micro-gap) between abut-  Figure 3: Abutments of the Meoplant implant system
                                                              ment and implant and ensures better adhesion. The MeoClassic®